24 Verification
Table 2-7. Performance Test Record for Agilent Model 6571A or 6671A
MODEL Agilent_____________
Report No.______________
Date_____________________
Test Description
Minimum
Spec.
Results
*
Maximum
Spec.
Constant Voltage Tests
Voltage Programming
and Readback
Low Voltage (0V) V
out
Front Panel Display Readback
-8mV
V
out
- 12mV
________mV
________mV
+8mV
V
out
+ 12mV
High Voltage (8V) V
out
Front Panel Display Readback
7.9888V
V
out
- 16mV
_________V
_______mV
8.0112V
V
out
+ 16mV
Load Effect
V
out
- 0.46mV _______mV V
out
+ 0.46mV
Source Effect
V
out
- 0.46mV _______mV V
out
+ 0.46mV
PARD (Ripple and Noise)
Peak-to-Peak
RMS
0
0
_______mV
_______μV
7mV
650μV
Transient Response Time
(at 900 μs)
0 _______mV 100mV
Constant Current Tests
Current Programming
and Readback
Low Current (0A) I
out
Front Panel Display Readback
-125mA
I
out
- 150mA
_______mA
_______mA
+125mA
I
out
+ 150mA
High Current (220A) I
out
Front Panel Display Readback
219.655A
I
out
- 370mA
_________A
_______mA
+220.345A
I
out
+ 370mA
Current Sink (10A) Display
Readback
I
sink
- 160mA
________mA
I
sink
+ 160mA
PARD (Ripple and Noise)
RMS
0 ________mA
200mA
Load Effect
I
out
- 21mA ________mA
I
out
+ 21mA
Source Effect
I
out
- 21mA ________mA
I
out
+ 21mA
*Enter your test results in this column.