14 7820A GC Site Preparation Guide
1 7820A GC Site Preparation
Environmental Conditions
Operating the instrument within the recommended ranges
optimizes instrument performance and lifetime. Performance
can be affected by sources of heat and cold from heating, air
conditioning systems, or drafts. See Table 5. The conditions
assume a noncondensing, noncorrosive atmosphere.
Table 5 Environmental conditions for operation and storage
Product Conditions Operating temp
range
Operating
humidity range
Maximum
altitude
Agilent 7820A Standard oven ramp 5 to 45 °C 5 to 90% 3,100 m
Storage –20 to 65 °C 0 to 90%