Installation 
Cary 100/300/4000/5000/6000i/7000 Spectrophotometers User’s Guide  35 
Nitrogen purging 
Cary 100/300 
The sample compartment of the Cary 100/300 can be nitrogen purged 
if the Extended Sample Compartment is fitted. The nitrogen supply 
should be attached to the inlet tubes beneath the Extended Sample 
Compartment. 
Refer to the Extended Sample Compartment operating instructions 
in the Help for further information. 
Cary 4000/5000/6000i/7000 
The nitrogen purging system is NOT supplied by Agilent, but the 
following items are available from appropriate commercial suppliers: 
Nitrogen 
Ultra-high purity liquid nitrogen, for example Nitrogen 5.0, (in 
conjunction with a heat exchanger) is recommended because it is less 
expensive than compressed nitrogen and is usually of better quality. 
If compressed nitrogen must be used, the gas must be dry, oil-free 
and uncontaminated.  
CAUTION 
Never use compressed nitrogen from a supplier who uses oil or water in the 
compression process. Such processes invariably leave fine particles of oil 
suspended in the nitrogen, which will form an oily deposit on optical surfaces. 
Use nitrogen from a supplier who fills containers from immersion pumps 
lubricated with liquid nitrogen. 
 
NOTE 
The instrument warranty will be void if damage is caused by the use of 
unsatisfactory nitrogen.