Chapter 3 Calibration Procedures
Test Considerations
43
3
Test Considerations
To ensure proper instrument operation, verify that you have selected the
correct power-line voltage prior to attempting any test procedure in this
chapter. See page 24 in chapter 2 for more information.
Assure that the calibration ambient temperature is stable and between 20°C
and 30°C.
Assure ambient relative humidity is less than 80%.
Allow a 1-hour warm-up period before verification or calibration.
Keep cables as short as possible, consistent with the impedance
requirements.
Caution The tests should be performed by qualified personnel. During performance
verification tests, hazardous voltages may be present at the outputs of the power
supply.