Chapter 1 Specifications
Performance Specifications
15
1
Performance Specifications
Output Ratings (@ 0°C - 40°C)
Low range 0 to +15 V/0 to 7 A
High range 0 to +30 V/0 to 4 A
Programming Accuracy
[1]
12 months (@ 25°C ± 5°C), ±(% of output + offset)
Voltage 0.05% + 10 mV
Current 0.2% + 10 mA
Readback Accuracy
[1]
12 months (over GPIB and RS-232 or front panel with respect
to actual output @ 25°C ± 5°C), ±(% of output + offset)
Voltage 0.05% + 5 mV
Current 0.15 % + 5 mA
Ripple and Noise (with outputs ungrounded, or with either output terminal grounded,
20 Hz to 20 MHz)
Normal mode voltage <0.35 mV rms and 2 mV p-p
Current <2 mA rms
Common mode current <1.5 mA rms
Load Regulation, ±(% of output + offset)
Change in output voltage or current for any load change within ratings with remote
sensing connected.
Voltage <0.01% + 2 mV
Current <0.01% + 250 mA
Line Regulation, ±(% of output + offset)
Change in output voltage and current for any line change within ratings
Voltage <0.01% + 2 mV
Current <0.01% + 250 mA
[1]
Accuracy specifications are after an 1-hour warm-up with no load and calibration at
25°C

Table of Contents

Other manuals for Agilent Technologies E3632A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E3632A and is the answer not in the manual?

Agilent Technologies E3632A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE3632A
CategoryPower Supply
LanguageEnglish

Related product manuals