Chapter 3 Calibration Procedures
General Calibration/Adjustment Procedure
64
Then, the display shows the low voltage calibration point.
6
Read the DVM and change the low voltage value on the display to match
the measured voltage.
For example, if the DVM reading is 0.4500 V, adjust the voltage to 0.4500 V using
the knob and resolution selection keys.
7
Pressing the "Calibrate" key saves the change and selects the middle
voltage calibration point.
If the entered number is within an acceptable range, an "ENTERED" message
appears for one second. If the entered number is not correct, an "MIN VALUE"
"MAX VALUE" message appears for one second and the display shows the low
voltage calibration point again. The display now shows the middle voltage
calibration point.
8
Read the DVM and change the middle voltage value on the display to
match the measured voltage.
For example, if the DVM reads 14.995 V, adjust the voltage to 14.995 V using
the knob and arrow keys.
9
Pressing the "Calibrate" key saves the change and selects the high
voltage calibration point.
v lo +0.5000 v
v lo +0.4500 v
v Mi +15.000v
v Mi +14.995v
v hi 29.500 v
Calibrate
Calibrate

Table of Contents

Other manuals for Agilent Technologies E3632A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E3632A and is the answer not in the manual?

Agilent Technologies E3632A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE3632A
CategoryPower Supply
LanguageEnglish

Related product manuals