Chapter 3 Calibration Procedures
Calibration Program
73
3
continued
370 !
380 ! Perform the DAC error correction, voltage calibration and OVP calibration.
390 ! Alert the operator to hook up the connection before calibrating.
400 !
410 ! Alert operator to connect lead
420 PRINT TABXY(10,10),"*********************************************************"
430 PRINT TABXY(10,11)," Prepare for E3632A DAC DNL error correction and"
440 PRINT TABXY(10,12)," Voltage/OVP calibration. Connect the output to the DMM."
450 PRINT TABXY(10,13)," Observe Polarity!"
460 PRINT TABXY(10,14),"*********************************************************"
470 PRINT TABXY(10,16),"Press 'C' to Continue, 'I' to go to CURRENT calibration or"
480 PRINT TABXY(10,17),"'X' to eXit, then press 'Enter'"
490 Ch$="C"
500 INPUT Ch$
510 IF Ch$="X" OR Ch$="x" THEN GOTO 2250
520 IF Ch$="I" OR Ch$="i" THEN
530 CLEAR SCREEN
540 GOTO 1460
550 END IF
560 CLEAR SCREEN
570 PRINT TABXY(10,7),"BEGIN DAC ERROR CORRECTION"
580 WAIT 4
590 CLEAR SCREEN
600 OUTPUT @Pwrsupply;"OUTP ON" ! Turn on Power Supply output
610 OUTPUT @Pwrsupply;"CAL:DAC:ERROR" ! Perform DAC DNL error correction
620 WAIT 29 ! Allow DAC error correction to finish
630 OUTPUT @Pwrsupply;"OUTPUT OFF" ! Turn off Power Supply output
640 OUTPUT @Pwrsupply;"SYST:ERR?"
650 ENTER @Pwrsupply;Error$
660 !
670 ! Check to see if there is an error. If there is an error,
680 ! display the error and exit the program.
690 !
700 CLEAR SCREEN
710 IF Error$="+0,""No error""" THEN
720 PRINT "DAC DNL Error Correction completed for Power Supply "
730 ELSE
740 PRINT "ERROR:";Error$;"DAC DNL Error not corrected "
750 BEEP
760 GOTO 2250
770 END IF
780 PRINT TABXY(10,5),"DAC DNL ERROR CORRECTION COMPLETE"
790 PRINT TABXY(10,7),"BEGIN VOLTAGE CALIBRATION"
800 WAIT 4
810 OUTPUT @Pwrsupply;"OUTPUT ON"
820 CLEAR SCREEN
830 OUTPUT @Pwrsupply;"CAL:VOLT:LEV MIN" ! set output to minimum cal value
840 WAIT 2 ! allow output to settle
850 OUTPUT @Dmm;"MEAS:VOLT:DC?" ! measure output with Dmm and
860 ENTER @Dmm;Dmm_rdg ! store in variable Dmm_rdg
870 PRINT Dmm_rdg

Table of Contents

Other manuals for Agilent Technologies E3632A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E3632A and is the answer not in the manual?

Agilent Technologies E3632A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE3632A
CategoryPower Supply
LanguageEnglish

Related product manuals