Chapter 5 Service
Troubleshooting Hints
97
5
Bias Supplies Problems
Check that the input to the voltage regulators of the bias supplies is at least 1
V greater than their output.
Circuit failures can cause heavy loads of the bias supplies which may pull down
the regulator output voltages.
Check the voltages of bias supplies as tabulated below.
Table 5-1. Bias Supplies Voltages
Some circuits produce their own local bias supplies from the main bias
supplies. Be sure to check that these local bias supplies are active. In particular,
the ADC (analog-to-digital converter), ac input, and front panel sections have
local bias supplies. Always check that the power supplies are free of ac
oscillations using an oscilloscope. Failure of bias supplies will cause many self-
test failures.
Bias Supply Minimum Maximum Check At
+5V Floating
-5.1V Floating
+15V Floating
-15V Floating
+4.75 V
-4.75 V
+14.25 V
-14.25 V
+5.25 V
-5.25 V
+15.75 V
-15.75 V
U11 pin 2
Anode of CR5
Anode of CR9
Cathode of CR10

Table of Contents

Other manuals for Agilent Technologies E3632A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E3632A and is the answer not in the manual?

Agilent Technologies E3632A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE3632A
CategoryPower Supply
LanguageEnglish

Related product manuals