Chapter 4 Remote Interface Reference
Status Reporting Commands
109
4
STATus:QUEStionable:ENABle <enable value>
This command enables bits in the Questionable Status enable register. The
selected bits are then reported to the Status Byte.
STATus:QUEStionable:ENABle?
This command queries the Questionable Status enable register. The power
supply returns a binary-weighted decimal representing the bits set in the enable
register.
*CLS
This command clears all event registers and Status Byte register.
*ESE<enable value>
This command enables bits in the Standard Event enable re gister. The selected
bits are then reported to the Status Byte.
*ESE?
This command queries the Standard Event enable register. The power supply
returns a decimal value which corresponds to the binary-weighted sum of all
bits in the register.
*ESR?
This command queries the Standard event register. The power supply returns
a decimal value which corresponds to the binary-weighted sum of all bits in
the register.
*OPC
This command sets the ‘‘Operation Complete’bit (bit 0) of the Standard Event
register after the command is executed.
*OPC?
This command returns ‘‘1’to the output buffer after the command is executed.
*PSC { 0 | 1 }
(Power-on status clear.) This command clears the Status Byte and the Standard
Event register enable masks when power is turned on (
*PSC 1
). When
*PSC
0
is in effect, the Status Byte and Standard Event register enable masks are
not cleared when power is turned on.