Chapter 4
Front-Panel Key and SCPI Command Reference
Measurement keys
205
Front-Panel Key and SCPI Command
Reference
4.4.4.5.6 Capture Time
Sets the capturing data time. When in manual mode, the user determines the capture time; otherwise, it is
determined automatically by the measurement.
This key is available only when PSA firmware revision A.11.08 or above, or Option 23A is installed.
Key Path: Meas Setup
Remote Command: [:SENSe]:AMDepth:TRIGger:SOURce
IMMediate|IF|EXTernal[1]|EXTernal[2]|RFBurst
[:SENSe]:AMDepth:TRIGger:SOURce?
Default: IMMediate
State Saved: Saved in instrument state.
Range: Free Run (Immediate)|Video (IF Envlp)|RF Burst|Ext Front|Ext Rear
Example: :AMD:TRIG:SOUR IF
:AMD:TRIG:SOUR?
Key Path: Meas Setup
Remote Command: [:SENSe]:AMDepth:CAPTure <time>
[:SENSe]:AMDepth:CAPTure?
[:SENSe]:AMDepth:CAPTure:AUTO 0|1|ON|OFF
[:SENSe]:AMDepth:CAPTure:AUTO?
Preset/Default: 250 ms
State Saved: Saved in instrument state.
Min: 1 μs
Max: 10 s
Resolution 1 ms
Example: :AMD:CAPT 0.25
:AMD:CAPT?
:AMD:CAPT:AUTO OFF
:AMD:CAPT:AUTO?

Table of Contents

Other manuals for Agilent Technologies E4440A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4440A and is the answer not in the manual?

Agilent Technologies E4440A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4440A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals