Chapter 4
Front-Panel Key and SCPI Command Reference
Measurement keys
279
Front-Panel Key and SCPI Command
Reference
4.4.14.2.16 TRFL Range
Returns the Range setting for the current TRFL measurement.The returned values are as follows:
1 = Range 1
2 = Range 2
3 = Range 3
4.4.14.2.17 Store and Recall all Instrument States Related to Cal Factors
.The following commands are used to set and query all instrument states related to cal factors.
Remote Command: [:SENSe]:TTRFlevel:RANGe?
Preset: 1
Range: 1 | 2| 3
Remote Command
Notes:
Query Only
Example: :SENS:TTRF:RANG?
Remote Command: :CALCulate:TTRFlevel:CALF para1, para2, ...para11
:CALCulate:TTRFlevel:CALF?
Remote Command
Notes:
SCPI query: returns the following 12 parameters:
Center Frequency (float64)
Input Attenuator (Integer32)
External RF Gain (Float64)
Cal Factor1 (Float64)
Cal Factor2 (Float64)
Cal Factor3 (Float64)
Range1-Range2 switching level (float64)
Range2-Range3 switching level (float64)
Track Range (float64Integ)
Integ BW (float64)
Accuracy (Enum, NORM/HIGH)
TunedLevel (it is mixer level + input attenuator, float64)
SCPI set command: sets the cal factor values. There must be 12 parameters attached to the
SCPI command. The 12 parameters can be retrieved by using the query command. The
returned values can then be copied to the parameter string.