Chapter 2 55
Installation and Setup
Calibrating System Components
Installation and Setup
Press System, More, More, Power Sensor Config, More, Update Power-On
Default
to save the current Sensor and Cal Factors information in the PSA and
will be loaded next time the PSA is powered on.
To recall the power sensor calibration factors from the Flash memory of PSA.
Press
System, More, More, Power Sensor Config, Cal Factors, More, Recall
Power-On Default
to load the Cal Factors file saved in the PSA.
The format of the .XML File is shown in the following example:
<Calibration>
<ModelNumber>N5532A</ModelNumber>
<Options>518</Options>
<SerialNumber>US44330015</SerialNumber>
<DateCalibrated>20000101</DateCalibrated>
<ReferenceCF>98</ReferenceCF>
<CalFactor>
<Frequency Unit= “Mhz”>10</Frequency>
<CalFactor Unit= "Percent">99.2135310306974</CalFactor>
</CalFactor>
<CalFactor>
<Frequency Unit= “Mhz”>30</Frequency>
<CalFactor Unit= “Percent”>98.7875312060724</CalFactor>
</CalFactor>
</Calibration>
Most of the SCPI commands for the calibration factors file operations are
consistent with those in the MMEMory subsystem, for example, Catalog, Delete,
etc., except for Save and Load.
File Operations for Tuned RF Level Calibration Factors
TRFL measurements can support store and recall functions for the convenience of
repetitive TRFL measurements at many different frequencies. Because the cal
factors are dependent on the instrument setting and TRFL measurement setting, cal
factors are saved into a state file with instrument and measurement settings. When
you recall the cal factors, not only the cal factors are recalled, but all of the
instrument and measurement settings are recalled.
NOTE Cal factor values will become stale due to time and temperature drift, and
following the PSA align routines.
From the front panel
To start one set of cal factors at one frequency:
1. Press
File, Save, Type, State, and enter a file name for this set of cal factors.
2. Press
Save Now.

Table of Contents

Other manuals for Agilent Technologies E4440A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E4440A and is the answer not in the manual?

Agilent Technologies E4440A Specifications

General IconGeneral
BrandAgilent Technologies
ModelE4440A
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals