104 Chapter 2
Instrument Functions: A - L
Display
Instrument Functions: A - L
Range: −3 kHz to 350 GHz; −30 Gsec to 30 Gsec
History: Added with firmware revision A.03.00
2.5.5.1.5.3 Amplitude
Pressing
Amplitude allows you to enter the amplitude value for the current limit point. After entering a value,
Connected becomes active. If a front-panel arrow key is pressed without entering a value, the current Amplitude
and
Connected values of the point are selected. If the up arrow is pressed, the point number automatically
increments to allow entry of the amplitude of the next point. If it is a new point, the
Frequency may also be
entered for the new point.
Key Path:
Display, Limits, Limit 1, Edit
Display, Limits, Limit 2, Edit
Factory Preset
and *RST: 0 dBm
Fundamental
Units: dBm
Terminators: dBm
Default Terminator: dBm
Resolution: 0.1 dBm
Knob Increment: 0.1 dBm
Step Key
Increment: 10 dBm
Range: -140 dBm to 100 dBm
History: Added with firmware revision A.03.00
2.5.5.1.5.4 Connected To Previous Pt
A current point may be connectd to the previous point by pressing Yes. No limit testing is performed between
disconnected points. Pressing this key when the Connected field is selected toggles the connected value of the
current point and increments the Point number to allow entry or editing of the Frequency of the next point. If
an arrow key is pressed without entering a value, the current Connected value of the point is selected. If an
arrow key is pressed, the Point number automatically increments to allow entry of the Connected value of the
next point. If it is a new point, the
Frequency may be entered for the new point
Key Path: Display, Limits, Limit 1, Edit
Display, Limits, Limit 2, Edit
Factory Preset
and *RST: Yes
Knob Increment: Toggles to other selection.
Step Key
Increment: Toggles to other selection
History: Added with firmware revision A.03.00