170 Chapter 3
Instrument Functions: M - O
Marker
Instrument Functions: M - O
3.1.3 Delta
Sets the control mode for the selected marker to Delta (see Marker on page 167). In Delta
mode the display shows the difference between the active (
Delta) marker and a reference
marker. When
Delta mode is selected the reference marker is placed at the current marker
position. If the marker is
OFF both the active marker and the reference marker are placed
at the center of the display. The amplitude and frequency (or time) of the first marker is
fixed. The marker number is indicated above the delta marker, and the same number is
indicated with an R (for example, 1R) above the reference marker. You can adjust the trace
point of the active delta marker. Annotation in the active function block and in the
upper-right corner of the display indicates the frequency or time difference and amplitude
difference of the two markers. If marker noise is set to On while using Delta and the noise
marker is placed on the noise floor, the marker readout displays signal to noise.
Selecting
Delta while already in Delta mode causes the reference marker to be reset to the
current active (
) marker position, enabling you to make delta measurements from
differing reference points without having to turn off the markers and begin again. Pressing
Marker Normal moves the Reference Marker to the Delta Marker position and turns off
Delta Marker.
The amplitude of the reference marker is fixed. In non-zero spans the frequency of the
reference marker is fixed. If the center frequency of the analyzer is changed such that the
reference marker is off the screen, an arrow will appear with the marker number at the
left or the right side of the display. This indicates where the trace point is for the reference
marker.
In
Zero Span the reference marker remains fixed at the trace point on which it was placed.
Also, changing
Center Frequency does not move the reference marker while in Zero Span.
The markers will be turned off if the scale type is changed between log and linear.
Key Path:
Marker
Remote Command:
See Marker on page 167 for the mode command.
Example: CALC:MARK4:MODE DELT selects marker 4 as a delta marker and places a
reference marker at the marker 4 position. If marker 4 is OFF it places
both the active and the reference markers at the center of the display.

Table of Contents

Other manuals for Agilent Technologies E444xAU Option H37

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E444xAU Option H37 and is the answer not in the manual?

Agilent Technologies E444xAU Option H37 Specifications

General IconGeneral
BrandAgilent Technologies
ModelE444xAU Option H37
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals