Chapter 2 163
Instrument Functions: A - L
Input/Output
Instrument Functions: A - L
Remote Command:
[:SENSe]:MIXer:HARMonic:AUTO OFF|ON|0|1 or <boolean>
[:SENSe]:MIXer:HARMonic:AUTO?
[:SENSe]:MIXer:HARMonic <integer>
[:SENSe]:MIXer:HARMonic?
Example: MIX:HARM:AUTO 0
MIX:HARM 8
2.8.9.2 Mixer Type
Selects which type of mixer is in use.
Mixer Type (Presel) activates a tuning signal that is
routed to the
PRESEL TUNE OUTPUT connector on the rear panel of the analyzer. This
signal has a sensitivity of 1.5V/GHz of the LO frequency and drives the tune input of the
HP/Agilent 11974 series of preselected mixers. The sweep rate in
Presel mode is limited to
40 MHz/ms.
Key Path:
Input/Output, External Mixer, Mixer Config
Dependencies/
Couplings: Only available in bands A, Q, U and V, and only when
Mixer Bias is off.
State Saved: Saved in Instrument State
Factory Preset: Unpre
Remote Command:
:INPut:MIXer:TYPE PRESelected|UNPReselect
:INPut:MIXer:TYPE?
Example: INP:MIX:TYPE UNPR

Table of Contents

Other manuals for Agilent Technologies E444xAU Option H37

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E444xAU Option H37 and is the answer not in the manual?

Agilent Technologies E444xAU Option H37 Specifications

General IconGeneral
BrandAgilent Technologies
ModelE444xAU Option H37
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals