EasyManuals Logo
Home>Agilent Technologies>Measuring Instruments>E444xAU Option H37

Agilent Technologies E444xAU Option H37 User Manual

Agilent Technologies E444xAU Option H37
434 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #51 background imageLoading...
Page #51 background image
Chapter 2 51
Instrument Functions: A - L
AMPLITUDE / Y Scale
Instrument Functions: A - L
2.1.11.2.2 Edit
Accesses menu keys that allow you to create and edit an amplitude-correction factor set. It
puts the analyzer into a split-screen mode where the correction data is displayed in a table
under the trace data. Pressing
ESC while in this menu will exit the menu and remove the
table from the screen. New points will be applied only after the editor is closed.
Key Path:
AMPLITUDE / Y Scale, More, Corrections, Antenna (Cable, Other, or User)
Remote Command:
[:SENSe]:CORRection:CSET[1]|2|3|4:DATA
<freq>,<rel_ampl>{,<freq>,<rel_ampl>} Creates an amplitude-correction factor set
[:SENSe]:CORRection:CSET[1]|2|3|4:DATA:MERGe
<freq>,<rel_ampl>{,<freq>,<rel_ampl>} Adds the points with the specified values to
the current amplitude correction data, allowing you to merge correction data. If too much
data is merged, as many points as possible are merged into the existing data and then an
error is reported.
[:SENSe]:CORRection:CSET[1]|2|3|4:DATA?
Remote Command Notes: [:SENSe]:CORRection:CSET:ALL[:STATe] must be set to on
for this command to function.
CSET number equivalents to front-panel access definitions are as follows:
CSET or CSET1 is Antenna
CSET2 is Cable
CSET3 is Other
CSET4 is User
<freq> is the frequency (in Hz) where the correction should be applied;
no unit is allowed in this parameter
<rel_ampl> is the amount of relative amplitude correction (in dB)
needed; no unit is allowed in this parameter
Example: :CORR:CSET2:DATA 900E6,0.3,1.0E9,0.35,1.3E9,0.2

Table of Contents

Other manuals for Agilent Technologies E444xAU Option H37

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E444xAU Option H37 and is the answer not in the manual?

Agilent Technologies E444xAU Option H37 Specifications

General IconGeneral
BrandAgilent Technologies
ModelE444xAU Option H37
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals