Chapter 2 99
Instrument Functions: A - L
Display
Instrument Functions: A - L
Remote Command:
:CALCulate:LLINe[1]|2:TYPE UPPer|LOWer
:CALCulate:LLINe[1]|2:TYPE?
Example: :CALC:LLIN2:TYPE LOW sets limit line 2 as a lower limit.
:CALC:LLIN1:TYPE? responds with the limit line 1 limit type.
2.5.5.1.2 Limit Display
Turns limit-line display
On or Off. Either Limit or Margin, as well as Test, must be turned on to
turn on a limit test.
Key Path:
Display, Limits, Limit 1
Display
, Limits, Limit 2
State Saved: Not saved in instrument state.
SCPI Status Bits/
OPC Dependencies: No OPC dependencies.
Factory Preset
and *RST: Off
Factory
Default: Off (when a limit line is created)
History: Added with firmware revision A.03.00
Remote Command:
:CALCulate:LLINe[1]|2:DISPlay OFF|ON|0|1
:CALCulate:LLINe[1]|2:DISPlay? queries the current limit line.
Example: :CALC:LLIN2:DISP OFF turns off the display of the limit lines.
:CALC:LLIN1:DISP? tells you whether the limit lines are being displayed.

Table of Contents

Other manuals for Agilent Technologies E444xAU Option H37

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies E444xAU Option H37 and is the answer not in the manual?

Agilent Technologies E444xAU Option H37 Specifications

General IconGeneral
BrandAgilent Technologies
ModelE444xAU Option H37
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals