110
Repairing the Autosampler
Needle Assembly
6 Reconnect the sample-loop fitting to the
needle fitting.
7 Use Needle Up to lift the needle to a
position approximately 2mm above the seat.
8 Ensure the needle is aligned with the seat. If
required, bend the needle slightly until the
needle is aligned correctly.
On completion of this procedure:
Install the front cover.
Select End in the maintenance function
Change Needle (see Change Needle on
page 81).