Repairing the Autosampler
Needle Assembly
6 Reconnect the sample-loop fitting to the
needle fitting.
7 Use “Needle Up” to lift the needle to a
position approximately 2mm above the seat.
8 Ensure the needle is aligned with the seat. If
required, bend the needle slightly until the
needle is aligned correctly.
On completion of this procedure:
❏ Install the front cover.
❏ Select “End” in the maintenance function
“Change Needle” (see “Change Needle” on
page 81).