106 Agilent InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps User Manual
6
Error Information
Module Error Messages
Module Error Messages
These errors are pump specific.
Solvent Zero Counter
Error ID: 2055, 2524
Pump firmware version A.02.32 and higher allow to set solvent bottle fillings
in the data system. If the volume level in the bottle falls below the specified
value the error message appears when the feature is configured accordingly.
Probable cause Suggested actions
1
Volume in bottle below specified volume.
Refill bottles and reset solvent counters.
2
Incorrect setting.
Make sure the limits are set correctly.