EasyManuals Logo
Home>Agilent Technologies>Water Pump>G7111B

Agilent Technologies G7111B User Manual

Agilent Technologies G7111B
262 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #146 background imageLoading...
Page #146 background image
146 Agilent InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps User Manual
8
Maintenance
Exchange the Purge Valve Frit or the Purge Valve
Exchange the Purge Valve Frit or the Purge Valve
For bio-inert modules use bio-inert parts only!
When Frit when piston seals are exchanged or when contaminated or blocked (pressure drop of
> 10 bar across the frit at a flow rate of 5 mL/min of water with purge valve opened)
Purge valve if internally leaking
Tools required p/n Description
8710-0510 Open-end wrench 1/4 — 5/16 inch
8710-1924 Open-end wrench 14 mm
Pair of tweezers
OR Toothpick
Parts required # p/n Description
1 G7111-60061 Purge valve
for G7110B, G7111A, and G7111B
1 G5611-60064 Purge valve
for G5654A (Bio-inert)
1 01018-22707 PTFE frits (pack of 5)
1 5067-4728 Seal cap (OPTIONAL)
Preparations Switch off the pump at the main power switch.
Open the doors of the module.
Use an optional solvent shutoff valve or lift up solvent filters in solvent reservoirs for avoiding
leakages.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7111B and is the answer not in the manual?

Agilent Technologies G7111B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7111B
CategoryWater Pump
LanguageEnglish

Related product manuals