Agilent InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps User Manual 175
Maintenance
8
Replacing the Module Firmware
Replacing the Module Firmware
To upgrade/downgrade the module’s firmware carry out the following
steps:
1 Download the required module firmware, the latest FW Update Tool and
the documentation from the Agilent web.
http://www.agilent.com/en-us/firmwareDownload?whid=69761
2 For loading the firmware into the module follow the instructions in the
documentation.
Module Specific Information
There is no specific information for this module.
When The installation of newer firmware might be necessary
• if a newer version solves problems of older versions or
• to keep all systems on the same (validated) revision.
The installation of older firmware might be necessary
• to keep all systems on the same (validated) revision or
• if a new module with newer firmware is added to a system or
• if third party control software requires a special version.
Tools required Description
Agilent Lab Advisor software
Parts required # Description
1 Firmware, tools and documentation from Agilent web site
Preparations Read update documentation provided with the Firmware Update Tool.