Agilent InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps User Manual 237
LAN Configuration
12
Initialization Mode Selection
Initialization Mode Selection
The following initialization (init) modes are selectable:
Default IP address for LAN is 192.168.254.11.
DHCP address is the module’s LAN MAC address.
Using Stored
When initialization mode Using Stored is selected, the parameters are taken
from the non-volatile memory of the module. The TCP/IP connection will be
established using these parameters. The parameters were configured
previously by one of the described methods.
Figure 25 Using Stored (Principle)
Table 2 6 Initialization Mode Switches
SW1 SW2 SW3 SW4 SW5 SW6 Init Mode
000000Use Default IP
Address
000010Use Stored IP Address
000100Use DHCP
Note: The setting ‘0’ (down) is essential.