Agilent InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps User Manual 125
Test Functions and Calibration
7
System Pressure Test
Running the Test
Running the test from the Agilent Lab Advisor
1 Select the System Pressure Test from the Test Selection menu.
2 Start the test and follow the instructions.
“Evaluating the Results” on page 126 describes the evaluation and
interpretation of the System Pressure Test results. For detailed instructions
refer to the Agilent Lab Advisor software.
Parts required p/n Description
5067-6127 Blank Nut SL
Damage to pressure sensitive parts
Even columns that are suitable for high pressures are sensitive to pressure drops that
occur during this test.
➔ Do not include any pressure sensitive parts to the flow path and choose a
maximum pressure that is compatible to your system. For example, do not include
columns, a standard pressure flow cell (up to 20 bar) or a 400 bar autosampler to a
600 bar pressure test.
Make sure to release the pressure by slowly opening the purge valve when the test has
finished.