62 Agilent InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps User Manual
3
Using the Pump
Setting up the Pump with the Instrument Control Interface
Instrument Configuration
Use the Instrument Configuration dialog box to examine and, if necessary, modify
your instrument configuration. The Configurable Modules panel contains a list
of all modules available for configuration. The Selected Modules panel contains
the list of configured modules.
Auto Configuration: Under Communication settings, select either the Host Name
option or the IP address option and enter the appropriate value for the host
computer to enable automatic detection of the hardware configuration. The
system configures the instrument automatically with no further manual
configuration necessary.
The Quaternary Pump configuration parameters are in two sections:
• Communication
•Options
Communication: The parameters in this dialog box are detected automatically
during autoconfiguration.
• Device name,
•Type ID,
• Serial number,
• Firmware revision,
• Button Connection settings
Options:
• Pressure Unit:
select the pressure units from the drop-down list (bar, psi or MPa).
• Seal wash installed:
This check box is marked to indicate that an optional seal wash has been
detected during autoconfiguration.
Please refer to the online help of your user interface for more detailed
information.