Error Information
Module Error Messages
Initialization Failed
Error ID: 2207, 2217
The module failed to initialize successfully within the maximum time window.
A maximum time is assigned for the complete pump-initialization cycle. If the
time is exceeded before initialization is complete, the error message is
generated.
Probable cause Suggested actions
1
Blocked (passive or active) inlet valve.
Exchange the (passive or active) inlet valve.
2
Defective pump drive assembly.
Please contact your Agilent service
representative.
3
Defective main board.
Please contact your Agilent service
representative.