EasyManuals Logo
Home>Agilent Technologies>Water Pump>G7111B

Agilent Technologies G7111B User Manual

Agilent Technologies G7111B
262 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #156 background imageLoading...
Page #156 background image
156 Agilent InfinityLab LC Series 1260 Infinity II Isocratic- and Quaternary Pumps User Manual
8
Maintenance
Maintenance of a Pump Head Without Seal Wash Option
3
Remove the support rings from the piston housing.
4
Lift the housing away from the pistons.
5
Check the piston surface and remove any deposits or
layers: clean the piston surface with abrasive paper and
rinse with 2-propanol. Replace piston if scratched.
6
Using the steel side of the insert tool, carefully remove
the seal from the pump housing.
7
Using the plastic side of the insert tool, insert new seals
into the pump head.
8
Place the support rings on the piston housing. Note the
correct position of the pins.
3LVWRQ6XUIDFH