EasyManuals Logo
Home>Agilent Technologies>Thermostat>G7116B

Agilent Technologies G7116B User Manual

Agilent Technologies G7116B
203 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #39 background imageLoading...
Page #39 background image
Agilent InfinityLab LC Series Multicolumn Thermostats User Manual 39
2 Site Requirements and Specifications
Valve Specifications
Table 8 G4231A (5067-4282), 2-Position/6-Port Valve 800 bar
Type Specification
Maximum pressure 800 bar
Typical application Any two-way switching, e.g. between two detectors, between
waste and detector, between two columns
Port size Accepts 10-32 male threaded fittings
Liquid contacts PEEK, Stainless steel
pH range
0 14
1
1
incompatible with some mineral acids. For more information see Solvent Information.
Table 9 G4232D (5067-4240), 2-Position/10-Port Valve 1300 bar
Type Specification
Maximum pressure 1300 bar
Typical application Anything a 2-Position/6-Port valve can do plus alternating column
regeneration
Port size Accepts 10-32 male threaded fittings
Liquid contacts PEEK, Stainless steel
pH range
0 14
1
1
incompatible with some mineral acids. For more information see Solvent Information.
Table 10 G4232C (5067-4283), 2-Position/10-Port Valve 800 bar
Type Specification
Maximum pressure 800 bar
Typical application Anything a 2-Position/6-Port valve can do plus alternating column
regeneration
Port size Accepts 10-32 male threaded fittings
Liquid contacts PEEK, Stainless steel
pH range
0 14
1
1
incompatible with some mineral acids. For more information see Solvent Information.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies G7116B and is the answer not in the manual?

Agilent Technologies G7116B Specifications

General IconGeneral
BrandAgilent Technologies
ModelG7116B
CategoryThermostat
LanguageEnglish

Related product manuals