4
Installation Procedure
Installation
44/102 HS452, HS652 Rotary Vane Pumps User Manual / 87-900-978-01 (H)
WARNING!
During installation, pay maximum attention that the suction flange is
connected to the vacuum chamber and the exhaust flange is not closed (see
the following figure). The pump must not be used as a compressor.
Maximum pressure inside the oil container must not exceed 1.5 bar (abs.).
Non-observance of these precautions may be dangerous for the machine and
the operator.
Figure 2
CAUTION!
Check that your electrical mains voltage corresponds to that indicated on the
motor’s plate.
Connect the pump to the power supply.