Installation Procedure
Installation
4
HS452, HS652 Rotary Vane Pumps User Manual / 87-900-978-01 (H) 43/102
Installation
Do not install or use the pump in an environment exposed to
atmospheric agents (rain, snow, ice), dust, aggressive gases, or in
explosive environments or those with a high fire risk. During
operation, the following environmental conditions must be respected:
temperature: from +12 °C to +40 °C
relative humidity: 0 – 95 % (non-condensing)
WARNING!
Protect against short circuits and overload by installing on Agilent Device
electrical main line an automatic circuit breaker of proper capacity (see
table here below):
Tab. 1
P/N 110 Vac 220 Vac
949-9360 10 A
949-9365 6 A
CAUTION!
Before starting the pump, fill up with lubricating oil as the pump is delivered
empty.
WARNING!
Take out the protective caps on the suction and exhaust flanges before doing
anything else. In the event of an accidental start-up, the air inside the pump
could violently expel the protective caps and harm the operator.