5
Technical Information
Pump Electronic Controller
80/102 HS452, HS652 Rotary Vane Pumps User Manual / 87-900-978-01 (H)
Pump Electronic Controller
Figure 9
Technical Specifications
Input voltage: 100-120 V/50-60 Hz; 200 V-240 V/50-60 Hz
Output voltage: 189 Vrms 3 ph; Max frequency: 68 Hz
(factory setting)
Starting phase maximum power: 780 W
(maximum time: 15 min.)
Normal Operation maximum power: 650 W
Maximum room temperature: 40 °C
Protection level: IP 44 (only using the accessory kit
P/N 949-9367)
CE mark:
EN55011 cat “B”
EN61000-3-2
EN61000-3-3
ESD EN61000-4-2
EN61000-4-3