18 Agilent Intuvo 9000 GC Installation
1 Installing the GC
Grounding
To protect users, the metal instrument panels and cabinet are
grounded through the three-conductor power line cord in
accordance with International Electrotechnical Commission
(IEC) requirements.
The three-conductor power line cord, when plugged into a
properly grounded receptacle, grounds the instrument and
minimizes shock hazard. A properly grounded receptacle is one
that is connected to a suitable earth ground. Proper receptacle
grounding should be verified.
Make sure the GC is connected to a dedicated receptacle
(outlet).
Brazil 230 10 NBR 14136
Type N
* Users of 100 V power (Japan) receive the 120 V US power option with a 100 V/120 V step up transformer.
Table 2 Power cords and terminations (continued)
Country
Line
Voltage
Amp
Rating
Cable length
(m)
Termination type
Plug
Termination
WARNING
Appliance Coupler (mains input power cord) is the power
disconnect device. Do not position the instrument such that
access to the coupler is impaired.

Table of Contents

Other manuals for Agilent Technologies Intuvo 900

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies Intuvo 900 and is the answer not in the manual?

Agilent Technologies Intuvo 900 Specifications

General IconGeneral
BrandAgilent Technologies
ModelIntuvo 900
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals