Installing the GC 1
Agilent Intuvo 9000 GC Installation 59
Connect the external cables.
Additional cables may be installed for control of the GC’s
automatic liquid sampler (ALS), connecting signal output to
integrators, synchronizing the start and end of a run between
various instruments, sensing conditions external to the GC, and
controlling devices external to the GC.
If using Event or BCD cables, label the cables as needed to
identify their intended use and appropriate connector on the
GC. See Labeling BCD and EVENT cables.
Back panel connectors
The figure below shows the connectors on the back panel of the
GC.
See also “Cable Diagrams” on page 106.
Sampler connectors
If using an ALS, connect it to the GC using the following
connectors:
ALS 1 Optional. An injector or 150-position sample tray.
ALS 2 Optional. An injector or 150-position sample tray.
The SIG (analog output) connectors
Optional. Use SIG1 and SIG2 for analog output signals.
ALS 1
ALS 2
SIG1
SIG2
REMOTE
EVENT
BCD
LAN