EasyManua.ls Logo

Agilent Technologies N6705 - Page 89

Agilent Technologies N6705
210 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Using the Source Functions 3
Model N6705 User’s Guide 89
Step Parameters
Parameter: Description:
Start Setting (I
0
or V
0
)
End Setting (I
1
or V
1
)
Delay (T
0
)
The setting before the step.
The setting after the step.
The delay after the trigger is received before
the step occurs.
From the remote interface:
The setting before the step:
ARB:CURR:STEP:STAR 0,(@1)
ARB:VOLT:STEP:STAR 0,(@1)
The setting after the step:
ARB:CURR:STEP:END 1,(@1)
ARB:VOLT:STEP:END 5,(@1)
The delay after the trigger is received before the step occurs:
ARB:CURR:STEP:STAR:TIM 0.01,(@1)
ARB:VOLT:STEP:STAR:TIM 0.01,(@1)

Table of Contents