Using the Source Functions 3
Model N6705 User’s Guide 95
Exponential Parameters
Parameter: Description:
Start Setting (I
0
or V
0
)
End Setting (I
1
or V
1
)
Delay (T
0
)
Time (T
1
)
Time Constant (T
C
)
The setting before the waveform.
The end setting of the waveform.
The delay after the trigger is received
but
before the waveform starts
.
Time for the amplitude to go from the start
setting to the end setting.
The time constant of the curve.
From the remote interface:
The setting before the waveform:
ARB:CURR:EXP:STAR 0,(@1)
ARB:VOLT:EXP:STAR 0,(@1)
The end setting of the waveform:
ARB:CURR:EXP:END 1,(@1)
ARB:VOLT:EXP:END 5,(@1)
The delay after the trigger is received but before the waveform starts:
ARB:CURR:EXP:STAR:TIM 0.25,(@1)
ARB:VOLT:EXP:STAR:TIM 0.25,(@1)
Time for the amplitude to go from the start setting to the end setting:
ARB:CURR:EXP:TIM 0.75,(@1)
ARB:VOLT:EXP:TIM 0.75,(@1)
The time constant of the curve:
ARB:CURR:EXP:TCON 0.1,(@1)
ARB:VOLT:EXP:TCON 0.1,(@1)