Chapter 3 177
Instrument Functions: M - O
Marker
Instrument Functions: M - O
3.1.8.3 Time
Sets the marker readout to Time, displaying the time interval between a normal marker
and the start of a sweep or the time of the delta marker relative to the reference marker.
Time is the default setting in zero spans. With a span of zero, the time value is the time
position relative to the start of the sweep. In a delta-marker mode it is the (sweep) time
interval between the two markers.
Key Path:
Marker, Readout
Remote Command:
See Readout on page 176 for this command.
Example: CALC:MARK2:X:READ TIME
3.1.8.4 Inverse Time
Sets the marker readout to Inverse Time, displaying the reciprocal of (sweep) time
between two markers. This function is only available when in both zero span and in a
delta-marker modes. If the markers are at the same x position, the time between them is 0
,
so the reciprocal of sweep time is infinitely large. The display will show a very large
number.
Key Path:
Marker, Readout
Remote Command:
See Readout on page 176 for this command.
Example: :CALC:MARK2:X:READ ITIM

Table of Contents

Other manuals for Agilent Technologies PSA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies PSA Series and is the answer not in the manual?

Agilent Technologies PSA Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelPSA Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals