Chapter 4 237
Instrument Functions: P - Z
SWEEP
Instrument Functions: P - Z
Remote Command:
[:SENSe]:SWEep:EGATe:POLarity NEGative|POSitive
[:SENSe]:SWEep:EGATe:POLarity NEGative|POSitive?
Example: SWE:EGAT:POL NEG
SWE:EGAT:POL?
4.8.5.3 Delay
Controls the length of time from the time the gate condition is satisfied until the gate is
enabled.
Key Path:
Sweep, Gate Setup
State Saved: Saved in instrument state
Factory Preset: 57.7 microsecond
Resolution: 100 ns
Range: 0.0
µs to 100 seconds
Remote Command:
[:SENSe]:SWEep:EGATe:DELay <time>
[:SENSe]:SWEep:EGATe:DELay?
Example: SWE:EGAT:DEL 100US
SWE:EGAT:DEL?
4.8.5.4 Length
Controls the length of time that the gate is enabled.
Key Path:
Sweep, Gate Setup
State Saved: Saved in instrument state
Factory Preset: 461.6
µs
Range: 10
µs to 0.5 s
Remote Command:
[:SENSe]:SWEep:EGATe:LENGth <time>
[:SENSe]:SWEep:EGATe:LENGth?
Example: SWE:EGAT:LENG 1ms
SWE:EGAT:LENG?

Table of Contents

Other manuals for Agilent Technologies PSA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies PSA Series and is the answer not in the manual?

Agilent Technologies PSA Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelPSA Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals