EasyManuals Logo

Agilent Technologies PSA Series User Manual

Agilent Technologies PSA Series
434 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #99 background imageLoading...
Page #99 background image
Chapter 2 99
Instrument Functions: A - L
Display
Instrument Functions: A - L
Remote Command:
:CALCulate:LLINe[1]|2:TYPE UPPer|LOWer
:CALCulate:LLINe[1]|2:TYPE?
Example: :CALC:LLIN2:TYPE LOW sets limit line 2 as a lower limit.
:CALC:LLIN1:TYPE? responds with the limit line 1 limit type.
2.5.5.1.2 Limit Display
Turns limit-line display
On or Off. Either Limit or Margin, as well as Test, must be turned on to
turn on a limit test.
Key Path:
Display, Limits, Limit 1
Display
, Limits, Limit 2
State Saved: Not saved in instrument state.
SCPI Status Bits/
OPC Dependencies: No OPC dependencies.
Factory Preset
and *RST: Off
Factory
Default: Off (when a limit line is created)
History: Added with firmware revision A.03.00
Remote Command:
:CALCulate:LLINe[1]|2:DISPlay OFF|ON|0|1
:CALCulate:LLINe[1]|2:DISPlay? queries the current limit line.
Example: :CALC:LLIN2:DISP OFF turns off the display of the limit lines.
:CALC:LLIN1:DISP? tells you whether the limit lines are being displayed.

Table of Contents

Other manuals for Agilent Technologies PSA Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies PSA Series and is the answer not in the manual?

Agilent Technologies PSA Series Specifications

General IconGeneral
BrandAgilent Technologies
ModelPSA Series
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals