XGS-600 Gauge Controller
26
DRAFT 4/
15/14
Soft Keys
SETPTS Select to open the Set Point screen (page 29).
SETUP Select to open the Sensor Setup screen (page 31) and other setup
screens.
LARGER Select to open the Large Font screen (page 40).
NOTE The NEXT key appears if there are more than eight gauges. Use this
key to move to the next Main screen with the additional gauges.
Sensor ID’s Lists the five-character User Label configured on the “Sensor Setup
Screen” on page 31.
Auto-On Indicator Indicates Auto-On has been set up, using the Sensor Setup screen,
between a convection and an ion gauge(s). The same symbol
appears next to both the convection and the ion gauge showing the
link between the two.
In the screen shown, convection gauge CNV2 has been set up to
trigger both ion gauges HFIG1 and HFIG2.
Multiple Auto-ON setups are possible, with a different symbol
shape for each Convection gauge.
See “Sensor Setup Screen” on page 31 for Auto-ON details.
Pressure Reading Displays the gauge pressure reading. The pressure units, as selected
in the System Setup screen, appear at the top of the column:
❑ Convection gauge readings always appear, as they are
always on.
❑ Active gauge readings always appear as they are always on.
The Active Gauge Card automatically recognizes the gauge
type and display pressure reading.
❑ For ion gauges, - - - appears when a gauge is first turned
on, until the software determines a valid reading.
See Table 4-1 on page 46 for error messages displayed in this
column.
If an ion gauge error message appears in the pressure reading
column and the control column indicates OFF, clear the error as
follows:
1. Press OK or reselect OFF in the control column.
2. Scroll to ON option to turn the gauge on.