EasyManuals Logo

Aim-TTI MX100T User Manual

Aim-TTI MX100T
46 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #38 background imageLoading...
Page #38 background image
Page 37
CONFIG <NRF> Sets the voltage tracking mode of the unit to <NRF> where <NRF>
has the following meaning:
0 = None.
1 = Mode1.
2 = Mode2.
3 = Mode3.
These modes are as defined within the Setting Voltage Tracking
section of this manualsee section 9.1.
CONFIG? Returns the voltage tracking mode of the unit. The response is
<
NR1><RMT>, where <NR1> has the following meaning:
0 = None.
1 = Mode1.
2 = Mode2.
3 = Mode3.
These modes are as defined within the Setting Voltage Tracking
section of this manual see section 9.1.
ONDELAY<
N> <NRF> Set output<N> Multi-On delay where <NRF> is in milliseconds
OFFDELAY<
N> <NRF> Set output<N> Multi-Off delay where <NRF> is in milliseconds
ONACTION<
N> <CPD> Set output<N> Multi-On action where <CPD> can be QUICK, NEVER
or DELAY.
OFFACTION<
N> <CPD> Set output<N> Multi-Off action where <CPD> can be QUICK, NEVER
or DELAY.
SAV<
N> <NRF> Save the current settings of output<N> to the store specified by
<
NRF> where <NRF> can be 0-49.
RCL<
N> <NRF> Recall the settings for output<N> from the store specified by <NRF>
where <
NRF> can be 0-49.
14.2.2 Common Commands
*IDN?
Returns the instrument identification.
The response is in the form <NAME>, <model>, <serial>, <version><RMT>
where <NAME> is the manufacturer's name, <model> is the instrument
type, <serial> is the interface serial number and <version> is the revision
level of the firmware installed.
*RST
Resets the functional parameters of the instrument to the default settings as
listed in the Factory Default Settings, see section 18.
Does not affect the contents of the Save and Recall stores.
Does not affect any remote interface settings.
*SAV <
NRF
>
Save the settings for all three outputs simultaneously to the store specified by
<NRF> where <NRF> can be 0-49. This includes output On/Off state, current
meter averaging state, and the Multi-On/Multi-Off settings.
*RCL <
NRF
>
Recall the settings for all three outputs simultaneously from the store specified
by <NRF> where <NRF> can be 0-49.This includes output On/Off state, current
meter averaging state, and the Multi-On/Multi-Off settings.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Aim-TTI MX100T and is the answer not in the manual?

Aim-TTI MX100T Specifications

General IconGeneral
BrandAim-TTI
ModelMX100T
CategoryPower Supply
LanguageEnglish

Related product manuals