17 Default Values
When supplied from the factory the power supply is set as follows:
All Outputs
Voltage 1 volt
Current 0.1 amp
Range 35V/3A
OVP 40V (O/P1 and O/P2); 80V (O/P3)
OCP 7A (O/P1 and O/P2); 3.5A (O/P3)
Current meter averaging Off (Level = Medium)
Voltage tracking None
Output On/Off Off
Multi On/Off action All Quick (synchronous on, synchronous off)
System Level Settings
Power-up state Settings as at power-down, all outputs off
Beep sound Enabled for key errors and trips
Disabled for entry into CC mode
Spin wheel action Normal
The default values can be restored from the Reset to factory Defaultsfunction see section