Page 38
Sets the Operation Complete bit (bit 0) in the Standard Event Status
Register. This will happen immediately the command is executed because
of the sequential nature of all operations.
Query Operation Complete status.
The response is always 1<RMT> and is available immediately the command
is executed because all commands are sequential.
Wait for Operation Complete true.
This command does nothing because all operations are sequential.
The product has no self-test capability and the response is always 0<
RMT
The product has no trigger capability. The command is ignored in this
instrument.
14.2.3 Status Commands
Clear Status. Clears all status indications, including the Status Byte.
Does not clear any Enable Registers.
LSR<N>? Query and clear the Limit Status Register<N>. The response format is
<
NR1><RMT>. See Status Reporting section for details of the response.
LSE<
N> <NRF> Set the Limit Status Enable Register<N> to <NRF>
LSE<
N>? Returns the value in the value in the Limit Status Enable Register<N>.
The response format is<
NR1><RMT>.
Query and clear Execution Error Register. The response format is
<NR1><RMT>.
Query and clear Query Error Register. The response format is
<NR1><RMT>.
Report the value of the Status Byte. The response is: <
NR
RMT
Because there is no output queue, MAV can only be read by a GPIB serial
poll, not by this query, as any previous message must have already been
sent.
NRF
Sets the Service Request Enable Register to <
NRF
Report the value in the Service Request Enable Register.
The response is <NR1><RMT>.
NRF
Set the Parallel Poll Enable Register to the value <
NRF
Report the value in the Parallel Poll Enable Register.
The response is <NR1><RMT>.
Returns the state of the ist local message as defined by IEEE Std. 488.2.
The response is 0<RMT> if the local message is false, or 1<RMT> if true.