27 \ USER OPERATION MANUAL: Purair Basic Series
6.8 Changing Out Filters
WARNING! Ensure persons removing lters are made aware of any potential hazards and that they are
provided with any necessary protective clothing and equipment.
Hazardsassociatedwiththeremovalanddisposalofusedlterswilldependontheapplicationofthehood.
Ifanactivatedcarbonlterisusedwithhydrocarbonsolvents,thelterwillretainthesolventswithoutloss,
andcanberemovedinthelaboratory.Theusedltershouldbesealedintoaplasticbagpriortodisposal,
preferably by incineration.
Ifthelterhascontainedanydangerousmaterialssuchasasbestosdustorradioactivechemicals,
operatorprotectionisadvised,includingtheuseofarespirator.Theusedltersmayrequiredisposal
by a specialist company.
NOTE: CONSULT YOUR SAFETY OFFICER OR INDUSTRIAL HYGIENIST BEFORE REMOVING OR
DISPOSING ANY FILTERS.
Pre-Filters
Thepre-lterislocatedbelowthemainlter.Removetheperforatedpre-ltertray.Removetheoldpre-lter
andplaceitintoabag.Sealfordisposal.Retthenewlterandretthepre-ltertray.
Main Carbon / HEPA Filter
WARNING! Disconnect the power supply before removing lter access cover.
• Removethefrontcovertogainaccesstothelter.Loosenthelterclamps.Liftthelterslightlytobreak
thesealandthenwithdrawthelter.Placethelterinaplasticbag.Sealthebagfordisposal.
• Slidethenewlterintopositionbypushingthelterfullyintothemodule.Retthefrontcoverandlockit
in position.
• Pleasenote,sometimesafternewltersaretted,itmaybenecessarytorecalibratetheairowsystem.
This procedure can be found in Calibration - page 26.
6.9AirowAdjustment
The speed controller on the cabinet can be accessed behind the main control panel.