16 \ USER OPERATION MANUAL: Purair Basic Series
2. Alignthetabsfortherearbafeinto
slotsinthetopbafekeepingthe1/4
turn clips facing down. This will support
therearportionofthetopbafeand
keeptherearbafeintheupright
position.
3. Securetopbafewithbracketsonthe
left, right and front sides.
4. Securethebafetothebracketsusing
1/4 turn clips. They are accessed from the
bottomsideofthebafe.
3.5PerformanceValidation/Certication
Afterinstallationandpriortouse,cabinetperformancemustbevalidatedandcertiedtofactorystandards.
The following tests should be performed:
AirowVelocity
Thetestingmethodsandequipmentrequiredarespeciedonthetestreport.Itisrecommendedthatthese
testsbeperformedonlybyaqualiedtechnicianwhoisfamiliarwiththemethodsandproceduresforcertify-
ing these types of cabinets.
3.6ImportanceofPerformanceValidation/Certication
Anairowvelocityvaluethatfallsbelowthevaluespeciedinsidethetestreportwillnotprovideadequate
operator protection.
3.7 Disclaimer
The performance of the cabinet, while rigorously evaluated at the factory, cannot be guaranteed after transit
and installation. Therefore on-site testing is always recommended.