23 \ USER OPERATION MANUAL: Purair Basic Series
4.2 Cabinet Operating Procedure
Thefumehoodshouldonlybeoperatedwiththecorrectlterinstalledfortheapplication.Referto
Filter Information - page 32 for further information. The ductless fume cabinet must not be used for labora-
toryworkinwhichchemicalsofdifferenttypesareusedthatdonotmatchtheltertypeorthattheprimary
chemicals and their byproducts are not known. The ductless fume cabinet should not be used for different
chemicalprocesseswherechemicalsfromthedifferentprocessescouldreactinthelter.
To start the system, apply power to the system and switch on the green power On/Off switch. The lights and
fan will automatically turn on.
Checktheairowandthelterconditionofthecabinetonaregularbasis.ThisiscoveredinMaintenance -
page 25.
Pleasenote,lterblocksdonotabsorbcarbonmonoxideorhydrogen.Smallquantitieswillnotcausehazards
because of the large dilution factor from the amount of air passing through the cabinet and the retardation of
thechemicalintheltermatrix.
Air Science fume hoods have been designed to handle fumes and vapors given off during everyday labora-
toryprocedures.Thesewillbeatthepartspermillion(PPM)levelintheairstreamenteringthelterblock.It
is not recommended that large quantities of solvents or acids be used or boiled off in the cabinet.
Intheeventofalargespillageinthecabinet,theamountoffumesenteringthelterblockmaytemporarily
reducetheefciencyofthelter.Forthisreasonanymajorspillagemustbeclearedupimmediately,prefera-
bly using spillage absorption granules rather than paper, which may aggravate the evaporation of toxic fumes
from the spillage area.
Followingamajorspillage,theltersmustbechanged,astheheatofwettingmayreducetheefciencyof
thelter.Afteraperiodofstabilization,theoldltersmaybereused,providingtheyhavenotreachedthe
saturation level.
The electrical equipment in the cabinet such as the lights and controls are not in the dirty air stream of the
system.Thesystemshouldnotbeusedinaammableroomatmosphere.Specialmodiedcabinetscanbe
provided for use in these areas. Contact Air Science for further information on these applications.
Operators should avoid sudden movements within the fume cabinet, such as rapid opening or closing of the
sashwindow,asthismaycausetemporaryreversaloftheairow.
The operators should maintain the normal safety equipment and procedures for dealing with hazardous chemicals.
WARNING!
» Donotuseagasame(Bunsenburners)wheneverpossible,asitinterfereswithairow.
» Do not change the original blower speed of the cabinet unless the change is required by a decrease in
measuredairvelocity.Adjustmentshouldbemadeonlybyaqualiedtechnician.Donotoperatethe
cabinet if fan fails to run.
» Minimizearmmovement.Movearmsinandoutofthecabinetslowlytoavoiddisruptingcabinetairow.
» Use absorbent pads on the work surface where appropriate to minimize splatter and aerosol generation
in case of spillage.
» Keep lids/covers on all containers, dishes or sample plates.