Troubleshooting
Problem Probable cause Solution
9. Image presented as mirror
image
Plate incorrectly placed in
barrier envelope
Place plate correctly.
The black side (printed) of the plate must be visible
through the transparent side of the barrier envelope.
The plate was placed in the
mouth of the patient the
wrong way round
The blue sensitive side of the plate must be facing
toward the tube.
10. Ghost images and shadows
on the X-ray exposure
Image data on the plate has
not been sufciently erased
Check the erasure settings
Inform your Service Technician
Replace the Erasure unit
Exposed plate subject to
effects of ambient light
Use barrier envelope or transfer box to keep
plate shielded from ambient light until scanning.
11. No image or faint image The X-ray dose used for the
plate was too low
Check X-ray parameters
Amplification level too low
(HV-value)
Increase the amplication (HV-value)
Wrong scanning mode used Choose a suitable scanning mode
Threshold setting too high Reduce the settings
Plate exposed to ambient light Use barrier envelope or transfer box to keep
plate shielded from ambient light until scanning.
12. Plate does not t into the
intake slot
Incorrect inlet guide (too small) Use Size #0 inlet guide for Size #0 plates. Use
Size #2 inlet guide for Size #1 and #2 plates.
Different manufacturer PSP Only use Air Techniques plates.
13. X-ray exposure has stripes
on image
Plate has been exposed Only use the plate within 24 hours of data erasure.
Part of the plate was exposed
to light during handling
Do not expose used plates to light. Exposure and
the scanning of the image data should be carried
out within half an hour.
Exposed plate subject to
effects of ambient light
Use barrier envelope or transfer box to keep
plate shielded from ambient light until scanning.
Plate was dirty or scratched Clean the plate with PSP cleaning wipes available
from Air Techniques.
Replace plates with scratches and call dealer for
maintenance service.
14 Pale stripes in image Too much stray light entered
during scanning
Darken the room
Rotate the ScanX Swift so that no direct light
falls on the lm entry slot
15. Blue LED illuminated
steady or blinking
No connection between
ScanX Swift and PC
USB-driver not installed or inactive
Check the USB cable
Activate ScanX Swift using software
16. Red status LED blinks Inlet guide is missing or not
correctly in position
Install inlet guide correctly.