Veb- MN (Excell-MCDSe 400)#A(15-5-2007)(in).doc
10
MONOLAR CURRENTS, ELECTRODES, ADJUSTMENT of the POWERS
CURRENTS for CUT
PURE CUT (PURE) – Cut without coagulating effect
This current is controlled by an automatic system (ADC System) for the self regulation of the functioning, which guarantees that the power
selected by the operators remains constant according to the characteristics of the tissues.
Control devices to be used (area C):
- With the selection key, select the current PURE, then adjust the power with the regulation keys;
- For the delivery (loud acoustic signal, and yellow Led), press on the yellow pedal of the twin foot-switch pedal.
Instructions for use, adjustments, and electrodes
Use the specific electrodes for the resectoscope from:
70-80W with wire loop electrodes, for soft tissues (i.e. into the bladder);
90-100W with wire loop electrodes, for strong tissues (i.e. TUR);
90-100W with ribbon loop (cutting/vaporization) electrodes, for soft tissues (i.e. into the bladder);
100-110W with ribbon loop (cutting/vaporization) electrodes, for strong tissues (i.e. TUR);
100-120W with vaporisation electrodes of small dimensions;
150-170W with vaporisation electrodes of big dimensions.
COAGULATING CUT (BLEND-1) – Cut with soft coagulating effect
This current is controlled by an automatic system (ADC System) for the self regulation of the functioning, which guarantees that the power
selected by the operators remains constant according to the characteristics of the tissues.
Control devices to be used (area C):
- With the selection key, select the current BLEND-1, then adjust the power with the regulation keys;
- For the delivery (loud acoustic signal, and yellow Led), press on the yellow pedal of the twin foot-switch pedal
Instructions for use, adjustments, and electrodes
As per the current PURE.
COAGULATING CUT (BLEND-2) – Not suitable at all.
COMBINATED CUT (CUT ALTERNATED TO COAGULATION) ( ENDO) – Not suitable at all.
CURRENTS for COAGULATION
“FULGURATION” COAGULATION AT HIGH VOLTAGE (FULG FORCED) - Strong superficial sparkling effect, and optimum
deep coagulating effect
This current guarantees a strong coagulating effect, both deep and superficial. It is suitable both for coagulations performed with
forceps/surgical instrument, and for coagulations performed by grazing the tissues with the active electrode.
This current is controlled by an automatic system (ADC System) for the self regulation of the functioning, which guarantees that the power
selected by the operators remains constant according to the characteristics of the tissues.
Control devices to be used (area D):
- With the selection key, select the current FULG FORCED, then adjust the power with the regulation keys;
- For the delivery (acute acoustic signal, and blue Led), press on the blue pedal of the twin foot-switch pedal.
Instructions for use, adjustments, and electrodes
Use the specific electrodes for the resectoscope from 60-70W.
“PIN POINT CONTACT” COAGULATION AT MEDIUM VOLTAGE (PIN POINT CONTACT) – Medium-Low superficial
sparkling effect, and optimum deep coagulating effect
This current guarantees a strong deep coagulating effect and a normal superficial coagulating effect. It is suitable both for coagulations
performed with forceps/surgical instrument, and for coagulations performed by grazing the tissues with the active electrode, just in case the
operators prefer a superficial effect which is less strong than the one of the FULG FORCED coagulation.
This current is controlled by an automatic system (ADC System) for the self regulation of the functioning, which guarantees that the power
selected by the operators remains constant according to the characteristics of the tissues.
Control devices to be used (area D):
- With the selection key, select the current PIN POINT CONTACT, then adjust the power with the regulation keys;
- For the delivery (acute acoustic signal, and blue Led), press on the blue pedal of the twin foot-switch pedal.
Instructions for use, adjustments, and electrodes
Use the specific electrodes for the resectoscope from 60-70W.
“SOFT” COAGULATION AT LOW VOLTAGE (SOFT) – Not suitable.
“SPRAY” COAGULATION AT VERY HIGH COLTAGE (SPRAY) – Very strong superficial sparkling effect, and good deep
coagulating effect
This current guarantees a good deep coagulating effect, and a very strong superficial coagulating effect. It is suitable both for the
coagulations performed with forceps/surgical instrument, and for coagulations performed directly with the active electrode, even far away
from the tissues.
This current is controlled by an automatic system (ADC System) for the self regulation of the functioning, which guarantees that the power
selected by the operators remains constant according to the characteristics of the tissues.
Control devices to be used (area D):