- With the selection key, select the current SPRAY, then adjust the power with the regulation keys;
- For the delivery (acute acoustic signal, and blue Led), press on the blue pedal of the twin foot-switch pedal.
Instructions for use, adjustments, and electrodes
Use the specific electrodes for resectoscope from 60-70W.
ADVICE
For the Cut with wire loop electrodes, use as follows:
Current PURE, from 70-80W into the bladder or on soft tissues, for a cut without coagulating effect;
Current BLEND-1, from 70-80W into the bladder or on soft tissues, for a cut with soft coagulating effect;
Current PURE, from 90-100W on the prostate, for a cut without coagulating effect;
Current BLEND-1, from 90-100W on the prostate, for a cut with soft coagulating effect.
For the Cut with ribbon loop electrodes, use as follows:
Current PURE, from 90-100W into the bladder or on soft tissues, for a coagulating cut;
Current BLEND-1, from 90-100W into the bladder or on soft tissues, for a cut with stronger coagulating effect;
Current PURE, from 100-110W on the prostate, for a coagulating cut;
Current BLEND-1, from 100-110W on the prostate, for a cut with stronger coagulating effect.
NB: In order to optimize the functioning and obtain the perfect cut without any “sticking” of the tissues on the electrode, the above
mentioned powers maybe have to be changed through small variations of 10-15W each time.
For the Vaporisation with rolls, use as follows:
Current PURE, from 100-120W with small rolls, and from 150-170W with bigger rolls.
For the Coagulation, use as follows:
Current SPRAY from 40-50W. Optimum to coagulate both with cutting electrodes (loop, blade type) and coagulation electrodes, even
without any contact with the tissues;
Current FULG FORCED, from 60-70W, for a coagulation both with cutting and coagulation electrodes;.
Current PIN POINT CONTACT, from 60-70W, for a coagulation with the contact between the electrode and the tissues.