19
After the white calibration has been performed,
the instrument indicates that the green standard is
required.
Place the instrument on the green standard and
press the operate button.
Aftercalibrationhasbeennished,theinstrument
indicates the days until next calibration is required.
Green check:
Click on the green tile symbol to perform green tile
check.
Place the instrument on the green standard and
press the operate button.
Afterthegreencheckhasbeennished,the
instrument indicates the days left until the next
check is required.