M21-400 REV.1.1 DVPLUS DIGITAL VISCOMETER - OPERATIONAL MANUAL
Figure 2-6
If using the scrollable speed list, press the dropdown button in the Speed box to open the list of available
pre-set speeds . Scroll up or down through the list until you find the pre-set speed you want, then simply touch it
to select it.
Table 2-6.2 shows the available pre-set speed selections:
DVPlus Pre-Set Speed List (RPM)
0.0
0.3
0.5
0.6
1.0
1.5
2.0
2.5
3.0
4.0
5.0
6.0
10
12
20
30
50
60
100
Table 2-6.2
2.7 Full-Scale Range
The Full-scale Range (FSR) is a calculation of the maximum viscosity value that can be measured by a spindle and
speed combination when used with the specific spring torque of the DVPlus Viscometer. This value represents the
measured viscosity that will be displayed when the instrument’s % torque reading is at 100. The FSR will be displayed