M21-400 REV.1.1 DVPLUS DIGITAL VISCOMETER - OPERATIONAL MANUAL
2-12.10
If the Viscosity Wizard obtains a torque measurement between 20-80%, the DVPlus may adjust the speed in order
to get as close to 50% torque as possible. When the best torque is reached, the DVPlus will display the “Wizard:
Results” screen (see Figure 2-12.11). The Wizard: Results screen will provide the user with the optimal spindle and
speed for the sample tested. This screen will also give the user the run time that the wizard took to arrive at that
conclusion.
2-12.11
When Configure Test is pressed, the configure viscosity screen will be populated with the recommended Spindle and
Speed parameters from the Viscosity Wizard.